I-87015 RTD Input Module 7 channel

The I-87015 provides 7 channels for measuring 2 or 3 wire RTD of Platinum 100/1000, Nickel 120 as well as Copper 100/1000 types.

Analog Input

  • Input Channels : 7 Differential
  • Input Type : Pt, Ni, Cu
  • Wire Connection : 2/3 Wire RTD
  • Resolution : 16-bit
  • Sampling Rate : 12 Samples/ sec (Total)
  • Accuracy : +/- 0.05% of FSR
  • -3dB BandWidth : 15.7Hz
  • Zero Drift : +/- 0.5 uV/ °C
  • Span Drift : +/- 25 uV/ °C
  • Common Mode Rejection : 150dB min
  • Normal Mode Rejection : 100dB
  • Input Impedance > 1M ohms
  • Individually Configurable Channels : Yes
  • Open Wire Detection : Yes
  • 4KV ESD Protection : Yes, Contact for each terminal
  • Intra-module Isolation, Field to Logic : 3000VDC

RTD Temperature Ranges

Type Code RTD Type Temperature Range °C
20 Platinum 100, α= 0.00385 -100 ~ 100
21 Platinum 100, α= 0.00385 0 ~ 100
22 Platinum 100, α= 0.00385 0 ~ 200
23 Platinum 100, α= 0.00385 0 ~ 600
24 Platinum 100, α= 0.003916 -100 ~ 100
25 Platinum 100, α= 0.003916 0 ~ 100
26 Platinum 100, α= 0.003916 0 ~ 200
27 Platinum 100, α= 0.003916 0 ~ 600
28 Nickel 120 -80 ~ 100
29 Nickel 120 0 ~ 100
2A Platinum 1000, α= 0.00385 -200 ~ 600
2B Cu 100 at 0?C,α= 0.00421 -20 ~ 150
2C Cu 100 at 25?C,α= 0.00427 0 ~ 200
2D Cu 1000 at 0?C,α= 0.00421 -20 ~ 150
2E Platinum 100, α= 0.00385 -200 ~ 200
2F Platinum 100, α= 0.003916 -200 ~ 200
80 Platinum 100, α= 0.00385 -200 ~ 600
81 Platinum 100, α= 0.003916 -200 ~ 600

LED Display

  • 1 LED as Power/Communication Indicator
  • 14 LEDs as High/Low Alarm Signals

Power

  • Power Consumption : 1.0W

Environment

  • Operating Temperature : -25 to 75°C
  • Storage Temperature : -30 to 75°C
  • Humidity : 5 to 95%, Non-condensing

Dimensions

31mm x 81mm x 114mm (W x D x H)

£156.00
I-87015 RTD Input Module 7 channel